Saurabh Jain,Abhishek Pathak,Rajeev Kant Pandey.[J].Chin J Traumatol,2014,17(1):50-53. [doi] |
Retrieval of a broken guide wire transfixing the hip through the proximal femoral nail hole |
|
DOI: |
KeyWord: Hip fractures Complications Instrumentation |
FundProject: |
|
Hits: |
Download times: |
Abstract: |
Retrieval of a broken guide wire transfixing the acetabulum or with intrapelvic migration is challenging and frustrating for surgeons. We here present a case report on a method to remove a broken guide wire transfixing the acetabulum through the proximal hole of recon nail using a grasping forceps. This method is little invasive, easy, time-saving and without need for changing the inital
fixation. |